Rice Lake FC-i/Si Series, A&D Weighing User Manual

Page 73

Advertising
background image

Step

Parameters

Example 1

Example 2

Example 3

1

Ws

200kg

30.0kg

600kg

2

Ns

20,000

30,000

60,000

3

d’

2

3

6

4

d

2 (=0.02kg)

5 (=0.005kg)

10 (=0.1kg)

Ns x d

20,000 x 2

30,000 x 5

6,000 x 1

5

1/No (=d/Ns)

1/10,000

1/6,000

1/6,000

6. Calculate the voltage sensitivity “Es”.

Es = (As-Ao) x 5,000 x 1/No (µV)

[5,000 means excitation voltage 5V]

Ao: Load cell output at zero point (mV/V)
As: Load cell output at full capacity (mV/V)

7. Check the voltage sensitivity “Es”.

Es

t

0.5µV

Æ Calculated “Ns x d” is fixed. In example 3 in the step 5, a 600kg x

0.1kg scale will be obtained.

Es<0.5µV

Æ Change “d” for new resolution 1/No.

d=1

Æ 2

d=2

Æ 5

d=5

Ж 10 Ж 1 (dividing “Ns” by 10)

In example 1,

the new d=5: 1/No=d/Ns=5/20,000=1/4,000 (200kg x 0.05kg)

In example 2,

the new d=1 and Ns=3,000: 1/No=d/Ns=1/3,000 (30kg x
0.01kg)

Using the new 1/No, go to step 6 and repeat it until Es

t

0.5µV in step

7.

Remote Scale kg/lb relationship

USA Version ONLY

‰ When the capacity was set by “kg”:

Let capacity x minimum division = Ws x W

d

(kg)

When:

d = 1, then W

d

(lb) = W

d

(kg) x 2

d = 2, then W

d

(lb) = W

d

(kg) x 5/2

d = 5, then W

d

(lb) = W

d

(kg) x 2

And:

Ws (lb) = W

d

(kg) x No

Example: 15kg x 0.002kg, No = 15kg/0.002kg =7,500

W

d

(lb) = 0.002 x 5/2 = 0.005 (lb)

Ws (lb) = 0.005 (lb) x 7,500 = 37.5 (lb)

‰ When the capacity was set by “lb”:

Let capacity x minimum division = Ws x W

d

(lb)

When:

d = 1, then W

d

(kg) = W

d

(lb) x 1/2

d = 2, then W

d

(kg) = W

d

(lb) x 1/2

d = 5, then W

d

(kg) = W

d

(lb) x 2/5

And:

Ws (kg) = W

d

(lb) x No

Example: 30lb x 0.005lb, No = 30lb/0.005lb =6,000