Specifications for Remote Scale

The scale has the ability of driving up to 4 load cells (350

:) in a remote scale

(platform). Set for the capacity range 0.5kg to 5,000kg (1 lb to 10,000 lb).

‰ The scale has the ability of driving up to 4 load cells (350:) in a remote scale

(platform).

‰ Capacity range for the remote scale is 0.5kg to 5,000kg (1 lb to 10,000 lb).

‰ Minimum output at zero point is 1mV.

‰ Maximum output at full load is 14mV.

Excitation Voltage of the scale is 5V.

‰ Cable length should be kept under 5m (16.5 ft.) for higher accuracy.

‰ Pin connection (JM:NJC-207-PF):

The Load Cell and Input Sensitivity

The relationship between load cell and input sensitivity (X) for the scale is follows:

‰ Example

Load Cell Capacity

100kg

A

Rated Output

3mV/V

B

Min., Division of Display

0.01kg

D

5,000 x x

B

D

A

X”= µV

‰ When a single Load Cell is used,

the following formula should apply:

‰ System design will be satisfactory if “X” is

greater than 0.5µV. In the example above

X” = 1.5µV.

Capacity and Resolution

‰ The resolution of the remote scale is automatically determined during the calibration

procedure. The following is to enable you to calculate the resolution for a given capacity.

1. Decide the capacity value and assign it to “Ws”. Maximum 5 digits.

2. Get the maximum count “Ns” for the capacity. Ignore the decimal point and add “0”

to Ws until it is 5 digits.

3. Calculate: d’=Ns/10,000.

4. Decide the minimum division “d”.

d’=1

Æ d=1

2

t

d’>1

Æ d=2

5

t

d’>2

Æ d=5

d’>5

Æ d=10 This should be changed to d=1, dividing Ns by 10.

5. Now “Ns x d” and the resolution 1/No = d/Ns can be determined.