Rice Lake FC-i/Si Series, A&D Weighing User Manual

Page 65

Advertising
background image

Using OP-01 Bar Code Reader

1. Plug the modular connector of RS-232C cable

into the port on the bottom of bar code reader,
and twist the lock to the right to lock it.

2. Connect the other end of the RS-232C cable to

the RS-232C interface of the scale.

3. Plug in the AC adapter for OP-01. AC adapter

jack is located at the side of the RS-232C cable
connector.

Lock

RS-232C cable

4. Set the F-Function

f-06/07/08-01="5"

according to the interface you connected.
For example, to use Ch.1 (PORT 1), set
f-07-01="5".

AC adapter plug

5. Read the next bar codes step by step, then the

bar code reader will be set to work with the
scale.

This sets the RS-232C to be 9600 bps and 8 bits data / non parity. Set the
F-Function

f-06/07/08-04="2" and f-06/07/08-05="2". For example, to use Ch.1

(PORT 1), set

f-07-04="2" and f-07-05="2".

By this set-up, the FC-i / FC-Si can accept bar codes of EAN-13/8, JAN-13/8,
UPC-A/E, Codabar (NW-7), Interleaved 2 of 5 and Code 39.

The UPC-A bar code can not be used together with EAN/JAN-13.
Refer to "Note 1" at "Step 4".

Note 1: When using UPC-A bar code, skip "Step 4".

STEP

Set-up Bar Code

Definition

Step 1

Start of configuration

Step 2

Rest to default settings

Step 3

Select RS-232C interface

Step 4

Refer to Note 1

Force UPC-A to EAN-13

Step 5

Disable transmitting UPC-A
check digit