Rice Lake FC-i/Si Series, A&D Weighing User Manual

Page 56

Advertising
background image

‰ Store the comparator limits into a specified ID memory.

Command

M L

,

1 2 3

,

+ 1 0 1 0

,

+ 9 9 0 C L

F

Reply

AK

ID number: Maximum 6 digit
Upper limit:

Maximum 9 digit including sign

Lower limit:

Maximum 9 digit including sign

‰ Store the total count and number of additions into a specified ID memory.

Command

M A

,

1 2 3

,

+ 6 7 8 9

,

1 2 3 C L

F

Reply

AK

ID # Upper Limit Lowe Limit

ID # Total Count # of Additions

ID number: Maximum 6 digits
Total count:

Maximum 9 digits including sign

Number of addition: Maximum 8 digits

‰ Request the scale to reply with the F-Function settings.

Command

? F C

,

0 5 0 1 C L

F

Request the setting of

f-05-01.

Reply

F C

,

0 5 0 1

,

0 C L

F

f-05-01 = 0

Command

? F C C L

F

Request to reply with the setting of the next F-Function.

Reply

F C

,

0 5 0 2

,

0 C L

F

f-05-02 = 0

‰ Store the specified F-Function setting value

Command

F C

,

0 5 0 3

,

+ 1 2 3 4 C L

F

Set “1234” as the Upper limit.

F-# Setting value

F-# Setting value

F-Function number: 4 digits
Setting value:

Maximum 8 digits including sign.

Reply

AK

Command

O N C L

F

Start with new settings.

Reply

AK

Having finished the "FC" command, send the "ON" command to start the scale with
new settings. The scale replies <AK> (06H) and starts.

Advertising