Rice Lake FC-i/Si Series, A&D Weighing User Manual
Page 40
F-02-X ACAI Operation & Min. Unit Weight
ACAI Mode When Unit Weight Entered by Sample Pieces.
0
ACAI is disabled.
1
ACAI automatic operation.
2
ACAI manual mode (using the ENTER key).
ACAI Mode When Unit Weight Entered by Keyboard or ID
0
ACAI is disabled.
1
ACAI manual mode (using the ENTER key). This setting
works when
f-02-01 setting is not “0”.
2
ACAI automatic obeys
f-02-01 setting.
Minimum Unit Weight.
(1d=1 weighing display division)
The factory setting may be different for some countries.
0
1/5 d (FC-i series), 1/20 d (FC-Si series)
1
1/100 d (FC-i series), 1/400 d (FC-Si series)
F-03-X M+ Accumulation Function
M+ Accumulation – Automatic or Manual
0
Manual accumulation (by pressing the M+ key).
1
Automatic accumulation (Positive data only).
+ or – Count Data Acceptable. (Manual Accumulation)
0
Positive data only (5d and above).
1
Positive and negative data (5d and above or –5d or below).
f-02-01
f-02-02
f-02-03
f-03-01
f-03-02
+5d
M+
Not Accepted
f-03-02=0
+
0
-
+5d
-5d
M+
Not Accepted
f-03-02=1
+
0
-