‰ Best operating temperature is between 20°C~25°C / 68°F~77°F at about 50%~60%

relative humidity. There shouldn’t be large temperature fluctuations.

‰ The weighing room should be kept clean and dry.

‰ The weighing table must be of a solid construction.

‰ Corners of rooms are best as they are less prone to vibrations.

‰ Don’t install the scale near heaters or air conditioners.

‰ Don’t install the scale in direct sunshine.

‰ Try to ensure a stable AC power supply when using the AC adapter.

‰ Keep equipment containing magnets away from the scale.

‰ Warm up the scale before use or leave it on standby overnight.

‰ Ground the scale chassis for electrostatic discharge if the weighing conditions warrant.

Calibration

Calibration of the scale is required when the scale is initially installed, or if a remote
scale is added. Please refer to “9

. CALIBRATION” for more calibration information.