Data format for the scale data – Rice Lake FC-i/Si Series, A&D Weighing User Manual

Page 62

Advertising
background image

Example: #04

“EOT” of ASCII code

‰ Repeat data

The control codes $SP, $CR and $LF can be used with “

+ maximum 2 digit

number”. That code will be repeated the number of times designated.

Example: $LF 9

Repeat “$LF” 9 times.

$SP 12 Put 12 “Spaces”.

‰ Link mark “&”

If you will send more than 2 lines of data, attach “&” to the end of the first line. Then,
the scale considers the data to be continued.

“Space” or “,” will be used to separate these data. You can skip them, but you
cannot skip “,” after “PF”. You must start with “PF,”.

Data Format for the Scale Data

[

” in examples below shows “Space” (20H).

Parameters for the scale data will be replaced by the format below when the scale
sends them out.

Data has a fixed number of digits including a sign and a decimal point. The
insignificant zeros are replaced by “Space (20H)” (except the ID number).

$PC

[

[

[

[ + 1 2 3 4

[ P C

1234 pcs / 9 digit data + 3 digit unit

$WT

[

[ + 4

.

3 2 1 0

[ k

g

4.3210 kg / 9 digit data + 3 digit unit

$UW

+ 1

.

2 3 4 5 6 7

[

[ g

1.234567 g / 9 digit data + 3 digit unit

$TR

[

[ + 1

.

2 3 4 5

[ k

g

1.2345 kg / 9 digit data + 3 digit unit

$TL

[

[ + 9 9 9 9 9 9

[ P C

999999 pcs / 9 digit data + 3 digit unit

$AN

[

[

[

[ 1 2 3 4

1234 times / 8 digit data

$CD

0 0 0

1 2 3

ID Number 000123 / 6 digit data

$NM

Z

[ D

[ 5

.

6 V

[

[

[

[ 12 characters

$CP

O K

Result is “OK” / 2 characters

[

[

Result is not available.

$DT

0 3

/

1 8

/

2 0 0 3

The order of date obeys

f-11-01.

$TM

1 2

:

3 4

:

5 6 12(h) 34(m) 56(s)

Advertising