3 after use, 4 power supply – Rice Lake FX-i WP Series, A&D Weighing User Manual

Page 13

Advertising
background image

11

Prevent foreign materials, such as powder, liquid and metal, from invading the area around the

weighing pan.

The breeze break (FZ-i WP series and FX-120i WP / FX-200i WP / FX-300i WP ) is provided as an

accessory. An anti-static treatment has been applied to the breeze break components, but they may
be charged with static electricity when they are unpacked or when the humidity is low. If the weight
value is unstable even when there are no drafts or the balance has a problem with repeatability,
remove the breeze break. Or wipe the clear plates with a moistened cloth, use the optional AD-1683
DC static eliminator or apply an anti-static spray.

3-3 After Use

Do not allow the balance to be immersed in water. Even though the balance complies with IP code,

the balance will not withstand being completely immersed in water.

Avoid mechanical shock to the balance.

Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.

Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that is

moistened with warm water and a mild detergent.

The edge of the weighing pan is sharp. Use much care when cleaning the pan.

3-4 Power Supply

When the AC adapter is connected, the balance is in the standby mode if the standby indicator is on

(refer to “3-5 Display Symbols and Key Operation”). This is a normal state and does not harm the
balance. For accurate weighing, warm up the balance for at least 30 minutes before use.