7-2 Calibration Using the Internal mass

(One-Touch Calibration, only for FZ-iWP series)

This function calibrates the balance using the internal mass.
The only operation required is to press the CAL key

Operation

1 Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the

weighing pan.

2 After carefully installing the small breeze break,

press the CAL key.

3 The balance displays Calin and performs calibration using the internal mass. Do not allow

vibration or drafts to affect the balance.

4 The balance displays end after calibration. If the “GLP output (info)” parameter of the function

table is set to “1” or “2”, the balance displays glp and outputs “Calibration Report” using the
RS-232C interface or stores the data in memory. For details on the calibration report format, refer to
“10-2 GLP Report”.

5 The balance will automatically return to the weighing mode after calibration.

About the internal mass

The mass of the internal mass may change due to corrosion or other damage caused by the operating
environment, or due to aging. Check the internal mass periodically. Correct the internal mass value as
necessary. For details, refer to “Correcting the internal mass value” of an additional manual.
To maintain the weighing accuracy, perform the calibration using an external weight periodically, as
described below.