41

Zero tracking ( trc )

This function tracks zero point drift caused by changes in the environment and stabilizes the zero point.
When the weight data is only a few digits, turn the function off for accurate weighing.

Note
Digit, when used for the FX-i
WP series balance, indicates a unit of minimum weighing value.

trc 0

The tracking function is not used. Used for weighing a very light sample.

trc 1

The tracking function is used. Normal zero tracking.

trc 2

The tracking function is used. Strong zero tracking.

trc 3

The tracking function is used. Very strong zero tracking.

Display refresh rate ( 5pd )

Period to refresh the display. This parameter influences “Stream mode” along with “Baud rate” and “Data
output pause”

Note
This item is selected automatically when the response rate is changed.

Decimal point ( pnt )

The decimal point format can be selected.

Auto display-ON ( p-on )

When the AC adapter is connected, the display is automatically turned on in the weighing mode, without
the ON:OFF key operation. Used when the balance is built into an automated system. 30-minute warm
up is necessary for accurate weighing.

Auto display-OFF ( poff )

When the AC adapter is connected and no operation is performed (inactivity state) for 10 minutes, the
display is automatically turned off and the standby indicator is illuminated.