6 description of the item “data format, Interval output mode, A&d standard format 5if type 0 – Rice Lake FX-i WP Series, A&D Weighing User Manual

Page 45: Dp (dump print) format 5if type 1

Advertising
background image

43

Interval output mode

The weight data is periodically output.
When the PRINT key is pressed, the balance starts to output the weight data at a preset interval time.
When the PRINT key is pressed again, the balance stops outputting the weight data.

Required setting

dout prt

6

Interval output mode


dout int Interval

time

Example

“For outputting the weight data periodically."

Caution

The balance may not transmit the data completely at the specified interval times and baud rate.
Set the baud rate higher.

9-6 Description of the Item “Data Format”

A&D standard format

5if type 0

This format is used when the peripheral equipment can receive the A&D format. If an AD-8121B is used,
set the printer to MODE 1 or 2.

This format consists of fifteen characters excluding the terminator.
A header of two characters indicates the balance condition.
The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus sign is used.
The unit, consisting of three characters, follows the data.








DP (Dump print) format 5if type 1

This format is used when the peripheral equipment can not receive the A&D format. If an AD-8121B is
used, set the printer to MODE 3.

This format consists of sixteen characters excluding the terminator.
A header of two characters indicates the balance condition. No overload header is used.
The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is not

zero or overloaded.

The unit, consisting of three characters, follows the data.