3 percent mode (%), Selecting the percent mode, Storing the 100% reference mass – Rice Lake FX-i WP Series, A&D Weighing User Manual

Page 21: Reading the percentage

Advertising
background image

19

5-3 Percent Mode (%)

This is the mode to display the weight value in percentage compared to a 100% reference mass and is
used for target weighing or checking the sample variable.

Selecting the percent mode

1 Press

the

MODE

key to select

%

(percent mode).

Storing the 100% reference mass

2 Press

the

SAMPLE

key to enter the 100% reference mass

storing mode.

Even in the storing mode, pressing the

MODE

key will switch

to the next mode.

3 Place a container on the weighing pan, if necessary.

Press the

RE-ZERO

key to cancel the weight (tare). The

balance displays 100 0

%

.

4 Place the sample to be set as the 100% reference mass on the

pan or in the container.

5 Press the

PRINT

key to store the reference mass. The

balance displays 100.00

%

. (The decimal point position depends

on the reference value. The reference mass stored, even if the
AC adapter is removed, is maintained in non-volatile memory.)

Note
If the balance judges that the mass of the sample is too light to
be used as a reference, it displays
lo . Do not use the sample.

6 Remove the sample.

Reading the percentage

7 Place a sample to be compared to the reference mass on the

pan. The displayed percentage is based on the 100% reference
mass.
While the stabilization indicator is on, pressing the

PRINT

key

will output the weight value, using the RS-232C serial interface.

Note
Peripheral equipment, that is sold separately , such as a
printer or a personal computer is required.