Class Item

and

Parameter

Description

0

600

bps

1

1200 bps

2

2400 bps

3

4800 bps

4

9600 bps

bp5
Baud rate

5

19200 bps

0

7 bits, even

1

7 bits, odd

btpr
Data bit, parity bit

2

8 bits, none

0

CR LF

Crlf
Terminator

1

CR

CR: ASCII code 0Dh

LF: ASCII code 0Ah

0

A&D standard format

1

DP format

2

KF format

3

MT format

4

NU format

type
Data format

5

CSV format

Refer to “9-6 Description of the Item

“Data Format””.

0

No limit

t-Up
Timeout

1

1 second

Selects the wait time to receive a

command.

0

No output

5if
Serial
interface

erCd
AK, Error code

1

Output

AK:ASCII code 06h

mltMLT
Programmable-unit (Multi-unit)

Sets an arbitrary coefficient.

Available only when programmable-

unit mode is selected.

Unit
Unit

Refer to “4. WEIGHING UNITS”.

id
ID number

Refer to “10. ID NUMBER AND GLP REPORT”

0

Normal weighing mode

1

Capacity indicator

apf
Application function

2

Statistical calculation mode

0

Number of data, sum

1

Number of data, sum, maximum,
minimum, average, range
(maximum-minimum)

ap fnc
Application 5taf

Statistical function
mode output items

2

Number of data, sum, maximum,
minimum, average, range
(maximum-minimum), standard
deviation, coefficient of variation

:

Factory settings.

Caution

The balance may not transmit the data completely at the specified refresh rate, depending on the
baud rate.
When a refresh rate of 20 times/second is used, set the baud rate to 4800 bps or higher.