65

Display

Error code

Description

EC,

E00

Communications error

A protocol error occurred in communications.
Confirm the format, baud rate and parity.

EC, E01

Undefined command error

An undefined command was received.
Confirm the command.

EC,

E02

Not ready

A received command can not be processed.
e.g. The balance received a Q command, but not in the

weighing mode.

e.g. The balance received a Q command while processing

a RE-ZERO command.

Adjust the delay time to transmit a command.

EC,

E03

Timeout error

If the timeout parameter is set to “t-Up1”, the balance did not
receive the next character of a command within the time limit
of one second.
Confirm the communication.

EC,

E04

Excess characters error

The balance received excessive characters in a command.
Confirm the command.

EC,

E06

Format error

A command includes incorrect data.
e.g. The data is numerically incorrect.
Confirm the command.

EC,

E07

Parameter setting error

The received data exceeds the range that the balance can
accept.
Confirm the parameter range of the command.

Other errors

If the errors described above can not be released or other
errors are displayed, contact the local A&D dealer.



17-3 Asking For Repair

If the balance needs service or repair, contact the local A&D dealer.
The balance is a precision instrument. Use much care when handling the balance and observe the
following when transporting the balance.

Use the original packing material.
Remove the weighing pan from the main unit