7-3 Calibration Using an External Weight

This function calibrates the balance using an external weight.

Operation

1 Connect the AC adapter and warm up the balance for

at least 30 minutes with nothing on the pan.

2 Press and hold the

CAL

key until Calout is

displayed, and then release the key.

3 The

balance

displays

Cal 0 .

If you want to change the calibration weight (a list of
usable weights is shown on page 28),
press the

SAMPLE

key and proceed to step 4.

If you use the calibration weight value stored in the
balance, proceed to step 5.

4 Specify the calibration weight value as follows:

SAMPLE

key To switch the display condition to:

“All of the segments blinking”
(calibration weight selection mode) or
“The last two digits blinking” (value
adjustment mode).

RE-ZERO

key (To increase the value)

MODE

key

(To decrease the value)
To select the calibration weight or
adjust the value.

PRINT

key

To store the new weight value. Even
if the AC adapter is removed, the
data is maintained in non-volatile
memory.

CAL

key

To cancel the operation and return to

_

Cal 0 .



Select a

weight

To next page

Press and

hold the key

Release

the key

e.g. Calibration

weight value
1000.12 g