4 description of the class “environment, display, Condition, Stability band width ( 5t-b ) – Rice Lake FX-i WP Series, A&D Weighing User Manual

Page 42: Hold function ( hold ) (animal weighing mode)

Advertising
background image

40

9-4 Description of the Class “Environment, Display”

Condition (

Cond

)

Cond 0

This parameter is for sensitive response to the fluctuation of a weight value. Used for
powder target weighing, weighing a very light sample or when quick response weighing is
required.

After setting, the balance displays FAST.

Cond 2

This parameter is for stable weighing with slow response. Used to prevent a weight value
from drifting due to vibration or drafts.
After setting, the balance displays SLOW.

Note

With “Hold function (Hold)” set to “ON (1)”, this item is used to set the averaging time.

Stability band width ( 5t-b )

This item controls the width to regard a weight value as a stable value. When the fluctuation per second
is less than the parameter, the balance displays the stabilization indicator and outputs the data. The
parameter influences the “Auto print mode”

5t-b 0

This parameter is for sensitive response of the stabilization indicator. Used for exact
weighing.

5t-b 2

This parameter ignores slight fluctuation of a weight value. Used to prevent a weight value
from drifting due to vibration or drafts.
Note
With “Hold function (
Hold)” set to “ON (1)”, this item is used to set the stabilization
range.

Hold function ( Hold ) (Animal weighing mode)

This function is used to weigh a moving object such as an animal.
When the weight data is over the weighing range from zero and the display fluctuation is within the
stabilization range for a fixed period of averaging time, the processing indicator illuminates and the
balance displays the average weight of the animal. When the animal is removed from the weighing pan,
the display returns to zero automatically.
This function is available only when the hold function parameter is set to “1” (the animal mode indicator

ANIMAL illuminates) and any weighing unit other than the counting mode is selected.

The stabilization range and averaging time are set in “Condition (Cond)” and “Stability band width (5t-b)”.

Weighing range

FZ-120 i WP
FZ-200 i WP
FZ-300 i WP
FX-120 i WP
FX-200 i WP
FX-300 i WP

0.200 g or over

FZ-1200 i WP
FZ-2000 i WP
FZ-3000 i WP
FX-1200 i WP
FX-2000 i WP
FX-3000 i WP

2.00 g or over

Averaging time

Stabilization range

Cond 0

5t-b 0

Cond 1

5t-b 1

Cond 2

2 seconds

Faster

4 seconds

8 seconds More accurate

5t-b 2

Lesser

Greater

Advertising