35

Operation

1 Press

the

ON:OFF

key to turn off the display.

2 While pressing and holding the

PRINT

key and the

SAMPLE

key, press the

ON:OFF

key. The balance displays p5 .

3 Press

the

PRINT

key. Then the balance displays the function switches.

4 Set the switches using the following keys.

SAMPLE

key

To select the switch to change the parameter.

RE-ZERO

key

To change the parameter of the switch selected.

0: To inhibit changes.

1: To permit changes

PRINT

key

To store the new parameter and return to the weighing mode.

CAL

key

To cancel the operation ( Clr is displayed.) To return to the weighing
mode, press the

CAL

key once again.

8-2 Initializing the Balance

This function returns the following parameters to factory settings.

Calibration

data

Function

table

The sample unit mass value (counting mode), 100%

reference mass value (percent mode)

External calibration weight
Function switch settings
Statistical

data


Note
Be sure to calibrate the balance after initialization.

Operation

1 Press the

ON:OFF

key to turn off the display.


2 While pressing and holding the

PRINT

key and the

SAMPLE

key, press the

ON:OFF

key.

The balance displays p5 .


3 Press

the

SAMPLE

key to display Clr

.


4 Press

the

PRINT

key.

To cancel this operation, press the

CAL

key.


5 Press

the

RE-ZERO

key.


6 Press

the

PRINT

key to initialize the balance.

The balance will automatically return to the weighing mode.

With these keys

held down

Press the key