31



5 Confirm that there is nothing on the pan and press the

PRINT

key. The balance measures the zero point.

Do not allow vibration or drafts to affect the balance.

The balance displays the calibration weight value.


6 Place the displayed calibration weight on the

pan and press the

PRINT

key. The balance

measures the calibration weight. Do not allow
vibration or drafts to affect the balance.


7 The balance displays end . Remove the

weight from the pan.


8 When the GLP output is set, the balance displays

glp and outputs “Calibration Report”.


9 The balance will automatically return to the

weighing mode.


10 Place the calibration weight on the pan and

confirm that the value displayed is within

±2 digits

of the specified value. If it is not within the range,
check the ambient conditions such as breeze and
vibration, also check the weighing pan. Then,
repeat steps 1 to 9.



Advertising