9-3 Details of the Function Table

Class Item

and

Parameter

Description

0

1

Cond
Condition

2

Fast response, sensitive value

FAST

MID.

Slow response, stable value

SLOW

With “Hold 1”, sets the averaging

time.

0

1

5t-b
Stability band width

2

Stable range is ±1 digit

Stable range is ±3 digits

The stabilization indicator illuminates

when the display fluctuation is

within the range. With “Hold 1”,

sets the stabilization range.

0

OFF

Hold
Hold function

1

ON

Holds the display when stable in animal

mode. With “Hold 1”, ANIMAL turns

on.

0

OFF

1

Normal

2

Strong

trc
Zero tracking

3

Very strong

Keeps zero display by tracking zero

drift.

0

5 times/second

1

10 times/second

5pd
Display refresh rate

2

20 times/second

Period to refresh the display

0

Point (.)

pnt
Decimal point

1

Comma (,)

Decimal point format

0

OFF

p-on
Auto display-ON

1

ON

Turns on the weighing mode display

when the AC adapter is connected.

0

OFF

poff
Auto display-OFF

1

ON (10 minutes)

Turns off the display after 10 minutes

of inactivity.

0

Displays

rng
Display at start

1

Does not display

Select whether or not to display the

minimum weighing value at weighing

start.

0

Does not sound

ba5fnc
Environment
Display

beep
Beep

1

Sounds

Select whether or not to sound the

beep when operating on keys.

0

No comparison

1

Comparison, excluding "near zero" when stable value or overloaded

2

Comparison, including "near zero" when stable value or overloaded

3

Continuous comparison, excluding "near zero"

Cp
Comparator mode

4

Continuous comparison, including "near zero"

0

OFF

bep-

LO buzzer

1

ON

0

OFF

bep-
OK buzzer

1

ON

0

OFF

Cp fnc
Comparator

bep-
HI buzzer

1

ON

Cp Hi
Upper limit
Cp lo
Lower limit

Refer to “9-9 Comparator Function”.

:

Factory settings. Digit is a unit of minimum weighing value