62

16. MAINTENANCE

In normal use, the balance can be cleaned with water. But, keep the following precautions so that

dust and water do not invade the balance.

Do not direct water pressure at the bottom of the balance.

Do not use powerful water jets.

Do not immerse the balance in water.

Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that is

moistened with warm water and a mild detergent.

Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or repair.
Use the original packing material for transportation.
While cleaning the balance and keeping it waterproof, cover the RS-232C interface connector using

the terminal cover or connect the waterproof RS-232C cable (AX-KO2737-500), and cover the AC
adapter jack. Also, insure that the underhook cap is in place.












If water accumulates on the waterproof diaphragm, a weight value may be difficult to become stable.

Clean the diaphragm while taking much care not to deform it.



AC adapter jack
When the AC adapter is not connected, cover
the jack.

RS-232C interface
Cover the interface connector or connect the
waterproof RS-232C cable (AX-KO2737-500).

Advertising