6. RESPONSE ADJUSTMENT

This function stabilizes the weight value, reducing the influence on weighing that is caused by drafts and
vibration at the place where the balance is installed.
The function has three stages as follows and can be changed by simple key operation.

Indicator Description

FAST

Fast response, but prone to drafts and vibration.
Good for target weighing.

MID.

SLOW

Slow response, but strong against drafts and vibration.
Good for weighing which requires a stable display.

Operation

1 Press and hold the

MODE

key until RESPONSE is displayed.

And then, release the key.

2 Press the

MODE

key to select a rate of the response

adjustment. Either FAST, MID. or SLOW can be selected.

3 After a few seconds of inactivity, or when the

PRINT

key is

pressed, the balance displays

e

end . Then, it returns to the

weighing mode and displays the updated response indicator. The

response indicator remains displayed for a while.


Note
Setting the response rate automatically changes the “Condition
(
Cond)” and “Display refresh rate (5pd)” parameters of “Environment,
Display (
ba5fnc)” in the function table, as shown below:

Indicator Cond (Condition)

5pd

(Display refresh rate)

FAST

0

2

MID.

1

0

SLOW

2

0


When the balance is to be used with other setting combinations, set
each parameter in the function table.