21. TERMS / INDEX

Terms

Stable value

The weight data when the stabilization indicator appears.


Environment

Ambient conditions such as vibration, drafts, temperature, static electricity and
magnetic fields which affect the weighing operation.


Calibration

Adjustment of the balance so that it can weigh accurately.


Output

To output the weight data using the RS-232C serial interface.


Zero point

A weighing reference point or the zero display. Usually refers to the value displayed
when nothing is on the weighing pan.


Digit

Unit of digital resolution. Used for the balance, a unit of minimum weighing value.


Tare

To cancel the weight of a container which is not included in the weight data.


Mode

Balance operational function.


Re-zero

To set the display to zero.


GLP

Good Laboratory Practice.


Repeatability

Variation in measured values obtained when the same weight is placed and removed
repetitively. Usually expressed as a standard deviation.

e.g. Standard deviation=1 digit: This means that measured values fall within

±1 digit

in the frequency of about 68%.


Stabilization time Time required after a sample being placed, until the stabilization indicator illuminates

and the weight data is displayed.


Sensitivity drift An affect that a change in temperature causes to the weight data. Expressed as

temperature coefficient.

e.g. Temperature coefficient = 2 ppm/

°C : If a load is 300 g and the temperature

changes by 10

°C, the value displayed changes by the following value.

0.0002%/

°C x 10°C x 300 g = 6 mg


IP65

IP code: Degree of protection provided by enclosures.
6: Dust-tight. No ingress of dust. 5: Protect against water jets.