Installation and Wiring

7

The standard connection is designed for 4-wire (non remote sensing) use. To convert to 6-wire (remote sensing)
applications, cut the two PC traces on either end of TB1 as shown in Figure 2-6.

Figure 2-6. Load Cell Wiring From Indicator

NOTE:

The load cell shield wire should be connected

to one of the load cell cable clamp screws located on
the load cell mating connector.

Shielding is connected at only one end
(typically at the indicator end). If
connected at the strain gauge end,
disregard Figure 2-7.

Figure 2-7. Load Cell Shield Wire Connection

2.9

Standard Surge Protection Board

The IQ 700 digital indicator comes with a factory installed surge suppression board. The suppression board stops
the flow of excess voltage to the CPU board and attaches to TB1 on the CPU board just by pressing it onto TB1 and
tightening the connector screws.

Figure 2-8. Surge Protection Diagram

J7

K1

S3

J5

J3

F1

J6

TB5

TB1

–EXC

–SEN

–SIG

+SIG

+SEN

+EXC

6

5

4

3

2

1

LOAD CELL CONNECTOR

1

2

U2

Q5

U18

+ Batt -

GND LO HI

TO POWER SUPPLY

Check load cell
color code for
proper wiring.

Traces

Traces

!

Caution

Six-Pin Female

CR3

CR4

CR2

CR1

E1

E2

E3

E4

E5

E6

E7

CHGND

+EXC

+SENSE

+SIG

–SIG

–SENSE

–EXC

COMPONENT SIDE

RIGHT SIDE

CIRCUIT SIDE