IQ 700 Installation Manual

4.4

Option 4 – Accumulator

The accumulate function is used to add weight data to a register for later access by the user. The accumulator can
keep a running total of weights entered, either automatically using setpoints, or manually pressing the

PRINT

key

when the accumulate function is active.

The IQ 700 has a ten digit accumulator available when Option 4 is enabled. To protect against multiple
accumulations of data, the accumulator option has a selectable return to

ZERO ACCUMULATION BAND

feature.

This feature ensures that the weight display must return to a value inside the zero band before the accumlator can be
performed.

Parameter 10 in the configuration mode be set for lb or kg base only. This avoids adding incorrect
data to the accumulator when switching between lb base and kg conversion. Clear the total
whenever the capacity is changed.

Table 4-6. Option 4 Configuration Parameters

Accumulation

Accumulators are activated in the following ways.

Activated by pressing the

PRINT

key.

Activated by the

PRINT COMMAND

in the auto batch mode (parameter 11

bAt2

).

If the data is valid (not in motion, overload, etc.) the display data is added to the content of the accumulator
and the display prompts with

PRINT

when configured for demand print or

Acc

if not in demand.

To prevent multiple accumulations of weight data, a

ZERO

threshold detection band (in display graduations),

requires the display data to return to zero before another accumulation/print can be performed.

Total Recall

To view the contents of the accumulation register on the display, use the following steps: press

TARE RECALL

which is redefined as

TOTAL

RECALL

key when option 4 is enabled. The tare recall LED flashes indicating the

instrument is in the total recall mode. While in total recall, the contents of the accumulator can be printed out when
the

PRINT

key is pressed. If the serial output is configured for continuous output, the total is continuously printed

out anytime the indicator is placed in total recall mode.

Total Reset

The total can be cleared to zero when in the

TOTAL RECALL

display mode by:

1. Press the

CE

key. The unit prompts with the display message

Sure

.

2. Press

CE

a second time to clear the contents of the accumulator resetting it back to zero. The display prompts

Clear

.

Refer to Section 4.12.3 on page 45 for individual serial command option parameters for Option 4.

Option

Prompt Display

Description

OP.4

OP.4 oF
OP.4 on

off
on – accumulator enabled

4.1 oFF

1
2
3
4
5
6
7
8
9

10
20
30
40
50

Zero band “off” data accumulator, anytime the

PRINT

key is pressed.

± 1 Display graduations
± 2 Display graduations
± 3 Display graduations
± 4 Display graduations
± 5 Display graduations
± 6 Display graduations
± 7 Display graduations
± 8 Display graduations
± 9 Display graduations
± 10 Display graduations
± 20 Display graduations
± 30 Display graduations
± 40 Display graduations
± 50 Display graduations

!