Options Configuration

21

4.0

Options Configuration

The IQ 700 offers a selection of optional features that
are available in the configuration of the indicator. The
options setup mode allows the operator to expand the
capabilities of the indicator. The available features are:

Option 1 - Expanded resolution

Option 2* - Analog output

Option 3 - Time and date

Option 4 - Accumulator

Option 5* - Local area network\tag\id\counter

Option 6 - Five-point linearization

Option 7 - Smart serial I/O

Option 8* - Analog output relay

Option 9 - Keyboard lockout

Option 10 - Setpoints

Option 11 - Remote Input

NOTE: *Options 2, 5, and 8 require additional
hardware support. Options 1 through 11 are currently
available in the standard EPROM KDA1921-1.

To access digital option configuration, close switches
SW1-2 and 3 (Figure 4-1). Press and hold the

ENT

key

until

OPtion

appears on the display. If the option mode

is not enabled (all options turned off), dashes appear
across the display (

- - - - -

).

Figure 4-1. Close Switches SW1-2 and 1-3 for Option

Configuration

In the option mode, primary function keys (

ZERO

,

GROSS/NET

, etc.) operate as secondary function keys

(Figure 4-2).

Figure 4-2. Front Panel Keys with Alternate Functions in

Options Menu

Use the

ZERO

key to select options 1 through 11 or the

TARE

RECALL

key to scroll backward through the

options. To enable a selected option, use the

GROSS/NET

key.

Some options, once enabled, may have an option
submenu available. Use the

TARE

key to access the

option submenu. To make parameter selections in the
option submenu, use the

lb/kg CONV

key. Display

prompting is provided in each case.

Key Functions Summary

Use the list below (Figure 4-3), as a quick reference
when selecting and configuring the options.

Figure 4-3. Quick Key Functions

DEAD LOAD (3)

CONF (2)

CAL (1)

3 4

2

1

OPEN

NORM

1

6

2

7

3

8

4

9

5

0

CE

ENT

TARE

RECALL

ZERO

NET

GROSS

TARE

lb/kg

CONV

PRINT

1 SET

POINT

2 SET

POINT

CAPACITY

5000 lb. X 0.5

ZERO

NET

GROSS

lb

kg

MOTION

M O D E L

700

1

6

2

7

3

8

4

9

5

0

CE

ENT

TARE

RECALL

ZERO

NET

GROSS

TARE

lb/kg

CONV

PRINT

1 SET

POINT

2 SET

POINT

CAPACITY

5000 lb. X 0.5

ZERO

NET

GROSS

lb

kg

MOTION

M O D E L

700

1 SET

POINT

1

2

4

5

ENABLE

DISABLE

OPTION

MENU

PARAMETER

SELECT

SELECT
OPTION

TARE

RECALL

1

6

2

7

3

8

4

9

5

0

CE

ENT

ZERO – Increments option number

NET/GROSS – Turns options ON/OFF

TARE – Steps through subparameters or functions as
ENTER for numeric changes

UNITS – Selects subparameters or digit position/decrements
digit for numeric changes

PRINT – Increments digit for numeric entries

TARE RECALL – Provides backstep in configuration mode

0 – 9 – Used to enter numeric values directly,
including ZERO/SPAN values in Option 2
or TIME/DATE in Option 3 mode

ZERO

NET

GROSS

TARE

lb/kg

CONV

PRINT