0 calibration, 1 zero calibration, 2 single slope span calibration – Rice Lake IQ 700 Configurable Weight Indicator User Manual

Page 50: 3 4 2 1 open, As sho wn in figure 5-1, Key. the display should read, Then return to a zero reading, The display will read, Zero, Figure 5-3. sw1 with all switches open

Advertising
background image

46

IQ 700 Installation Manual

5.0

Calibration

The IQ 700 indicator can be calibrated using single slope span calibration or five-point linearization. Zero must be
calibrated (see Section 5.1) before either span or linearization calibration can be performed.

Indicator option 2 (Section 4.2 on page 22) provides settings for zero, span, and trim adjustments. The settings are
all digital therefore no potentiometers are required. During the trim adjustments for zero/span, the analog output is
forced to the zero/span previously selected in option mode parameters 2.5 and 2.6. While reading the analog output,
the trim is increased or decreased from 0 to ± 175 until the reading corresponds with the values entered in
parameters 2.5 and 2.6.

5.1

Zero Calibration

Zero calibration is accomplished by the following steps:

1. Clear the scale (no load).

2. Close SW1-3 (dead load). The leftmost display digit should be flashing

C

as shown in Figure 5-1.

Figure 5-1. Close Switch 3

Figure 5-2. Close Switch 3 and Open Switch 1

3. Adjust dead load potentiometer to obtain a reading at or near zero (see Figure 5-2).

4. Open SW1-3 and close SW1-1 to put the indicator in the digital calibration mode. The display may change

from a zero reading, which is acceptable.

5. Press the

ZERO

key. The display should read

---CAL

, then return to a zero reading.

5.2

Single Slope Span Calibration

1. Place a test weight on the scale and wait for the motion LED to go out.

2. Enter weight value using keyboard. When correct, press

ENT

. The display will read

---CAL

briefly, then

return to the correct reading.

3. Recheck by clearing scale.

4. Open all switches to place the unit in normal weighing mode (Figure 5-3).

Figure 5-3. SW1 with All Switches Open

5. Close center bar and tighten center bar screws. Seal the unit if used in a legal-for-trade application.

DEAD LOAD (3)

CONF (2)

CAL (1)

3 4

2

1

OPEN

NORM

1

6

2

7

3

8

4

9

5

0

CE

ENT

TARE

RECALL

ZERO

NET

GROSS

TARE

lb/kg

CONV

PRINT

1 SET

POINT

2 SET

POINT

CAPACITY

5000 lb. X 0.5

ZERO

NET

GROSS

lb

kg

MOTION

M O D E L

700

FLASHING

DEAD LOAD (3)

CONF (2)

CAL (1)

3 4

2

1

OPEN

NORM

1

6

2

7

3

8

4

9

5

0

CE

ENT

TARE

RECALL

ZERO

NET

GROSS

TARE

lb/kg

CONV

PRINT

1 SET

POINT

2 SET

POINT

CAPACITY

5000 lb. X 0.5

ZERO

NET

GROSS

lb

kg

MOTION

M O D E L

700

FLASHING

3 4

2

1

OPEN

Advertising