IQ 700 Installation Manual

4.1

Option 1 – Expanded Resolution

Enabling option 1 increases the indicator resolution by allowing display graduation selections beyond the normal
20,000 in parameter 1 of the configuration mode.

Up to 80,000 displayed graduations are available when this option is on; however, applying such high gains to the
data may cause undesirable display instability in some applications.

The expanded resolution (OP.1) is shown in Table 4-1

and its menu structure in Figure 4-4

.

Table 4-1. Expanded Resolution Options

Option 1 Menu

Figure 4-4. Option 1 Menu

Refer to Section 4.12.3 on page 45 for individual serial command option parameters for Option 1.

Prompt Display

Interpretation

Expanded Resolution

OP.1 oF
OP.1 on

Off
On (expanded up to 80,000 displayed graduations)

XXXXXXX

XXXXXXX

XXXXXXX

XXXXXXX

OPTION 5

OPTION 4

OPTION 3

OPTION 2

CALIBR

CONFIG

OPTION 6

OPTION 7

OPTION 8

OPTION 9

OPTION 10

OFF

ON

EXPANDED

RESOLUTION

SERIAL

OPTION 1

OPTION 11