Example 3:
Enter or change values for verifiable configuration in single range
mode in the unit set under 1.7.x.

Select menu item ADC CON

)

Confirm menu item ADC CON

(if necessary press

and select the menu item VERIF.

) k p)

Verifiable configuration

)

Confirm menu item VERIF.

Accuracy class

(if necessary press ) ) () To confirm accuracy class 3/4

Range selection

)

Confirm menu item RANGES

(if necessary press k repeatedly) Select menu item SINGLE

Multi-interval scale

(

Confirm menu item SINGLE

Weights

)

Confirm menu item WEIGHTS

Verifiable scale interval

)

Confirm menu item E

( ) k p

Enter a value (e.g. 0.002 kg)

(if necessary press ) repeatedly)

until display E appears

k

Select menu item MAX.

Enter values for the following in
the same way:

Maximum capacity (e.g. 30 kg)

Continue as shown in example 1
after entering the maximum
capacity

max.

E

Weights

SINGle

Ranges

Class

Verif.

ADC-CON

Example 4:
Enter or change values for verifiable configuration in a multi-inter-
val scale in the unit set under 1.7.x.
(the same applies for multiple range mode).

Select menu item ADC CON

)

Confirm menu item ADC CON

(if necessary press
) k p)

and select the menu item VERIF.

Verifiable configuration

)

Confirm menu item VERIF.

Accuracy class

(if necessary press ) ) () To confirm accuracy class 3/4

Range selection

)

Confirm menu item RANGES

(if necessary press k repeatedly) Select menu item MULT. INT

Multi-interval scale

(

Confirm menu item MULT. INT

Weights

)

Confirm menu item WEIGHTS

Verifiable scale interval

)

Confirm menu item E

( ) k p

Enter a value (e.g. 0.002 kg)

(if necessary press ) repeatedly)

until display E appears

k

Select menu item RANGE 1
Enter values for the following in
the same way:

Range limit 1 (e.g. 6 kg)

Range limit 2 (e.g. 15 kg)

Maximum capacity (e.g. 30 kg)

Continue as shown in example 1
after entering the maximum
capacity

Range 1

E

Weights

Mult.Int

Mult.Int

Class

Verif.

ADC-CON