Key ) - > 2 sec Function
Allocation

Purpose
The key ) - > 2 sec is usually used
for the calibration/adjustment function.
The following additional functions can
be allocated to the key when the service
mode is activated:

– External linearization with default

weights (menu item 1.9.6)

– External linearization with the lineariza-

tion weights (menu item 1.9.7) entered
under menu item 1.18

– Setting the preload (menu item 1.9.8)
– Clearing the preload (menu item 1.9.9)

!

Once linearization has been completed,
or after a preload has been set or
cleared the function of the key )- > 2
sec must be reallocated back to its nor-
mal function in the Setup menu (e.g.
external calibration/adjustment with
default weights)

17

17

Menu structure for key ) - > 2 sec function allocation

1. 9.

Calibration, adjustment

1. 9. 1

External calibration/adjustment with default weights (service mode not required)

1. 9. 3

External calibration/adjustment with user-defined weights (entered under 1-18), (service mode not required)

1. 9. 6

External linearization with default weights

1. 9. 7

External linearization with user-defined weights (entered under 1-18)

1. 9. 8

Set preload

1. 9. 9

Clear preload

1. 9. 10

Key blocked

Advertising