Start-up

Unpacking

§ Unpack the device and check it immedi-

ately for any visible damage.

$ If you detect any damage, proceed as

directed in the chapter entitled “Care
and Maintenance“, under “Safety
Inspection“.

$ Save the box and all parts of the pack-

aging for any future transport. Unplug
all connected cables before packing the
equipment.

Check package contents

– Indicator
– Operating instructions (this document)
– Options (special accessories) as listed on

the bill of delivery; possible options are:
Real-time clock with battery back-up
Interface (RS-232, RS-485, analog
interface 4-20 mA, digital I/O)
Internal rechargeable battery
External rechargeable battery
24V module

Installation instructions
Avoid adverse influences at the place
of installation:

– Heat (heater or direct sunlight; opera-

tional temperature: -10

º

C to +40

º

C)

– Drafts from open windows and doors
– Extreme vibrations during weighing
– Aggressive chemical vapors
– Extreme moisture (according to IP pro-

tection class)

Start up the device

$ If necessary, acclimatize the device:

see next column

§ Connect the weighing platform and the

indicator, see page 7 (any type of
weighing platform or weighing cell that
meets the required specifications can be
connected to the indicator)

§ Establish a connection to the power

supply see the page after next

§ Warm up the device: see the page after

next (warm up time)

§ Configure the analog/digital converter

(ADC) see page 9

§ Carry out an alignment: For calibration

see page 21; for linearization see
page 20

Acclimatize the device
Condensation can form on the surfaces
of a cold device when it is brought into
a substantially warmer area. Therefore,
on transferring the device to a warmer
area make sure it is acclimatized for
about 2 hours at room temperature
(unplugged from power).