GMP-compliant Printouts
When the corresponding menu item is
active, the measured result is bracketed
on the printout by a GMP header and
a GMP footer (GMP = “Good Manufactur-
ing Practice").
The GMP header precedes the first
measured result. The GMP footer is printed
either after each measured result
(“ISO/GLP/GMP: For 1 application result,"
menu item

7.11.2

), or after the last result

in a series of measurements
(“ISO/GMP/GLP: For several application
results," menu item

7.11.3

). To end a

series of measured results, press and hold
the p key (> 2 seconds). In this case,
the A symbol is displayed after the
GMP header is printed and remains in the
display until the GMP footer is printed.

A GMP-compliant printout is generated
automatically at the conclusion of
calibration/adjustment routines, as well as
when you set or clear a preload.

If you use a label printer for GMP-
compliant printouts and menu item

7.11.3

is active, the header and footer are printed
on two different labels. To generate
GMP-compliant printouts on labels, select
menu item

7.11.2

.

Examples of GMP headers and
one example of a footer are shown
in the following.

Weighing platform WP1:
--------------------

Dotted line

14.01.2007

09:43

Date and time

1)

Typ

MW1P1

Midrics model

Ser.no.

12345678

Midrics serial no.

Vers.

1.1007.12.1

Software release for application

BVers.

01-25-01

Software release for basic version

--------------------

Dotted line

GMP footer:

--------------------

Dotted line

14.01.2007

09:45

Date and time

1)

Name:

Field for operator signature
Blank line

--------------------

Dotted line

1)

Interface YDO01M-332CLK (Option A31) required