Rice Lake Sartorius Midrics 1 and 2 Series - Operating Instructions User Manual

Page 66

Advertising
background image

Place empty container on
the platform

)

1. Tare the scale.

Note: If the automatic tare
function is enabled, you do
not need to press the ) key
to tare the scale; the tare
weight is saved automatically
when you place the container
on the platform.

2. Place 1

st

animal in container

20

3. Enter the number of

subweighing operations using
the keypad (in this example,
20 measurements)

r

4. Save the value entered and

begin averaging

The averaging routine does
not begin until the fluctuation
in weight value remains below
a defined threshold over three
consecutive measurements.
The number of subweighing
operations remaining is shown
in the numeric display.

Read the result of averaging

p

5. Print the results.

Note: If automatic printout
of results is enabled, you do
not need to press the p key;
the results are printed auto-
matically.

mDef

+

20

Configured printout:
see page 82

T

+

0.292 kg

x-Net +

0.183 kg

--------------------

When you unload the
weighing platform, the display
switches to the weight readout
automatically, unless con-
figured otherwise in the oper-
ating menu. The weighing
instrument is ready for the
next measurement.

66

Example:
Measuring the weight of one mouse.
Settings (changes in the factory settings required for this example):
Setup: Application: Animal weighing
Setup: PRTPROT (printout): 7.7.x (COM1) (see “Configuration” for options)

Advertising