Rice Lake Sartorius Midrics 1 and 2 Series - Operating Instructions User Manual

Page 59

Advertising
background image

Preparation

§ Switch on the scale: Press e
§ While all segments are lit,

press the ) key

§ Select the Application menu: Press k

repeatedly until

APPL

is displayed

§ Open the Application menu:

Press the ) key

§ Select the Counting application:

Press the k key repeatedly until the
desired menu item is displayed
and press ) to open the submenu

Application Parameters: Counting

3.5.

Minimum load for automatic
taring and automatic printing
3.5.1*

1 digit

3.5.2

2 digits

3.5.3

5 digits

3.5.4

10 digits

3.5.5

20 digits

3.5.6

50 digits

3.5.7

100 digits

3.5.8

200 digits

3.5.9

500 digits

3.5.10

1000 digits

3.6.

Minimum load for initialization
3.6.1*

1 digit

3.6.2

2 digits

3.6.3

5 digits

3.6.4

10 digits

3.6.5

20 digits

3.6.6

50 digits

3.6.7

100 digits

3.6.8

200 digits

3.6.9

500 digits

3.6.10

1000 digits

3.7.

Automatic taring: first weight tared
3.7.1*

Off

3.7.2

On

3.8.

Start application and load most
recent application data when the
Midrics is switched on
3.8.1

Automatic (on)

3.8.2*

Manual (off)

3.9.

Resolution for calculation
of reference value
3.9.1*

Display resolution

3.9.2

Display resolution
+1 decimal place

3.9.3 Display

resolution

+2 decimal places

3.9.4

Internal resolution

3.11

Parameter for saving weight values
3.11.1*

At stability

3.11.2

At increased stability

3.12. Average piece weight updating

3.12.1

Off

3.12.3*

Automatic

3.25. Tare function

3.25.1*

Add input value
(weight value) for
taring

3.25.2

Tare value can be
overwritten

* = Factory setting

§ Press ) to save your settings and (

(repeatedly) to exit the operating menu

Minimum Load
To tare container weights automati-
cally, set the minimum load in the
operating menu.

The minimum load required for
initialization of the weighing platform
is configured in the operating menu
under:

appl

Z:

3.6.

– The error code

Inf 29

is displayed

– The weighing platform is not initialized
– The preset reference sample quantity

is saved

You can choose from 10 settings,
ranging from

1 digit
to
1000 digits

Example: If the scale interval (d) is
1000 g and the minimum load is set
to 1000 digits (=1000 scale intervals),
a load of at least 1000 g is required
for autotaring.

Resolution for Calculation of
Reference Value
The resolution applied for calculating
the reference weight is defined in the
operating menu under:

appl

Z:

3.9.

The resolution for calculating the
reference sample weight is increased if
“+1 decimal place", “+2 decimal places"
or “With internal resolution" is selected.
With the “+1 decimal place" setting,
the net value is determined to one
additional decimal place (i.e., display
accuracy

+ 10); the “+2 decimal places"

increases display accuracy

+ 100, and

so on up to the maximum resolution
available.

Parameter for Saving Weight Values
The weight on the platform is saved
as a reference when the platform has
stabilized. If you select “At increased
stability," the average piece weight
stored will be more accurate and the
results more reproducible, but the
response time of the weighing instru-
ment might be longer.

Reference Sample Updating
The average piece weight (APW) is
updated automatically only when the
following 4 criteria are met:

1. The current piece count exceeds the

original piece count by at least two.

2. The current piece count is no more

than double the original piece count.

3. The new piece count is less than 1000.
4. The scale is stable in accordance with

the defined stability parameter.

Auto

Indicates that APW update is

active.

opt

Indicates that the reference sample

is currently being updated.
During an updating operation,

0pt

and

the updated piece count are displayed
briefly in the measured value line.

59

59

Advertising
This manual is related to the following products: