Calibration and adjustment, Setting or clearing the preload – Rice Lake Sartorius Midrics 1 and 2 Series - Operating Instructions User Manual

Page 53

Advertising
background image

Calibration and Adjustment

Purpose
Perform calibration to determine the dif-
ference between the value displayed and
the actual weight on the platform. Cali-
bration does not entail making
any changes within the weighing instru-
ment.

Perform adjustment to eliminate any dif-
ference determined, or to reduce
it to a level that is within the applicable
tolerance limits.

Configuration for Use in Legal Metrolo-
gy
To configure the Midrics for use in
legal metrology, adjust the switch on the
back of the display and control unit (A/D
converter configuration: Verif. selected
for legal-for-trade applications). The
switch is covered by a
protective cap.

Position:

§ Switch on the right: For use in

legal metrology

§ Switch on the left: External calibration/

adjustment accessible

Features
You can configure the parameters
listed below in the operating menu.
Which of the features listed here are
available depends on the connected
weighing platform.

– External calibration/adjustment blocked

in verified weighing instruments

– External calibration/adjustment with the

default weight value or standard weight
(not available on verified
instruments). Configure under:

Setup

wp-1

1.9.

: (Calibration and adjustment)

– Specify the weight for external

calibration/adjustment:

Setup

wp-1

1.18.

: (Enter calibration weight)

– Block the ) key to prevent use of the

two functions described above (1.9.10):

Setup

wp-1

1.9.

: (Calibration and adjustment)

– Calibrate first; then adjust

automatically or manually (not on
verified weighing instruments):

Setup

wp-1

1.10.

: (Calibration/adjustment sequence)

– Block external calibration/adjustment:

Setup

wp-1

1.16.

: (External calibration)

Note
On verified weighing instruments,
the external calibration/adjustment func-
tion is available only when the menu
access switch is in the “open" position,
which entails breaking the
verification seal. The equipment must be
re-verified after the seal has been broken.

Setting or clearing
the Preload

The preload can be set only when the
menu access switch is open (see
diagram).

§ The “set preload" function must be

assigned to the ) key (menu item
1.9.8).

§ After setting the preload, close the menu

access switch and reset the ) key to its
previous function (e.g., external calibra-
tion/adjustment with user-defined
weights) under menu item 1.9.

The preload can be cleared only when the
menu access switch is open.

§ The “clear preload" function must be

assigned to the ) key (menu item
1.9.9).

§ After clearing the preload, close the

menu access switch and reset the )
key to its previous function (e.g., external
calibration/adjustment with user-defined
weights) under menu item 1.9.

Preparation

§ Switch on the scale: Press e
§ While all segments are lit,

press the ) key

§ Select the Setup menu: Press k

repeatedly until

Setup

is displayed

§ Open the Setup menu:

Press the ) key

$ Select weighing platform 1, “

WP1

”:

Press the ) key, or

$ Select interface 1, “

COM1

" or interface

2, “

COM2

" (depending on the interface

used): Press the ) key

Setup

wp-1

1

1.9.

Calibration and
adjustment

1.9.1*

Ext. calibration/adjustment;
default weight

1.9.3

Ext. calibration/adjustment;
user-defined weight (menu
code 1.18.1)

1.9.8.

Set preload

1.9.9.

Clear preload

1.9.10

No function when you
press and hold ) > 2 sec

1.10. Calibration/adjustment

sequence

1.10.1

Calibration with automatic
adjustment

1.10.2* Calibration with adjustment

triggered manually

1.11. Zero-setting

range

1.11.1

1 percent/max. cap.

1.11.2* 2 percent/max. cap.

1.12.

Initial zero-setting range

1.12.2

2 percent/max. cap.

1.12.3

5 percent/max. load

1.12.4* 10 percent/max. load

1.13.

Tare/zero at power on

1.13.1* On
1.13.2

Off; load previous
tare value

1.13.3

Only zero at power on

1.16. External

calibration/

adjustment

1)

1.16.1* Accessible
1.16.2

2)

Blocked

1.17.

1.17.1

Grams

1.17.2* Kilograms
1.17.4

Pounds

1

)

1.18.

Enter calibration weight

1.18.1

External user-defined
weight; (enter value;
e.g.: 10,000 g)

§ Save settings and exit operating menu:

( key (repeatedly)

53

1

) = Setting cannot be changed on scales

verified for use in legal metrology

2

) = Factory setting on instruments verified

for use in legal metrology