Rice Lake Sartorius Midrics 1 and 2 Series - Operating Instructions User Manual

Page 22

Advertising
background image

22

Position the calibration/adjustment weight on the weighing platform

The difference between the measured value and the true weight of the sample will be
disayed with plus/minus signs.

External calibration

A printout will be generated if the calibration is not carried out and the procedure

Nom

+

10000 g

can be stopped by pressing the ( key.

Diff. -

2 g

--------------------

)

Activate calibration manually (press the ( key to stop calibration/adjustment)

The calibration weight is displayed once calibration is finished.

--------------------

A GMP-compliant printout is generated

14.01.2007

13:00

Type

MIS2

Ser.no.

12345678

Vers.

1.1007.12.1

BVers.

01-25-01

--------------------
External calibration
Nom

+

10000 g

Diff. -

2 g

External calibration
Diff. +

0 g

--------------------
14.01.2007

13:02

Name:

--------------------

Unload the weighing equipment