89

Error messages are shown in the main display. Error messages are shown permanently; Information messages are shown for 2 seconds.
After this the program automatically returns to the weighing mode.

Display

Cause Remedy

ERR101

Key is stuck

Release key or

Key pressed whilst switching on

Contact your customer service partner

ERR 320

Operating program memory faulty

Contact your customer service partner

ERR 340

Incorrect operating parameter (EEPROM)

Switch scale off and then on again,
if Err 340 is permanently displayed:
Contact your customer service partner

ERR 341

Loss of data

Contact your customer service partner

ERR 343

Loss of data in storage for

Contact your customer service partner

external Alibi memory transaction number

Inf 01

Data output not compatible with

Set output format correctly

output format

Inf 02

Calibration condition was not met

Do not carry out calibration until after 0 display

e.g. not tared or

Unload the weighing equipment

weighing pan loaded

Tare with the ) key

Inf 03

Calibration could not be completed

Keep to warm-up time and

within a certain time

calibrate again

Inf 07

Last operating function is not permissible

For setting changes please

use in legal metrology

Contact your customer service partner

Inf 08

Load on the scale is too heavy for scale to be

Please check if point 1.12 “Zero-setting

zeroed.

range” was adhered to in your configuration.

Inf 09

Taring not possible for gross zero

Zero the scale

Inf 10

Taring not possible with occupied tare memory

Taring only possible after deletion of the
application program

Inf 22

Reference takeover error,

Place a heavier weight on the platform

weight too small

Inf 23

Error in initializing an application

Contact your customer service partner

Inf 29

Minimum load not achieved

Reduce minimum load
(under Application, menu item 3.6)

Inf 71

Saving of measured value (or input) not possible

None

(e.g. control limit too small or too large)

Inf 72

Saving of measured value not possible

None

(e.g. maximum for item counter reached)

Inf 73

Stored data is deleted or cannot be read

Contact your customer service partner

Inf 74

Function is locked

None

(e.g. menu is locked)

Inf 98

No weighing platform connected

Contact your customer service partner

Inf 99

No weighing platform connected

Contact your customer service partner

NO WP

No weighing platform connected

Connecting weighing platform

Error Messages