Application: Averaging (Animal Weighing) V

With the Averaging application,
you can use your weighing platform
for calculating weights as the average
of a number of individual weighing
operations. This function is used to
determine weights under unstable
ambient conditions or for weighing
unstable samples (such as live animals).

Features

– Averaging starts manually or

automatically. Configuration:

appl

V:

3.18.

With manual start selected, the averag-
ing routine begins when you press
a key (provided the start conditions are
met. With automatic start selected,
averaging begins when you place the
first load on the platform (provided
the start conditions are met).

– Enter the number of subweighing

operations using the keypad

– Press the r key to select the desired

number of subweighing operations

– Activate info mode by pressing I

– Toggle the display between last result

and current weight by pressing w

– Automatic printout of results.

Configuration:

appl

V:

3.20.

– Automatic taring of container weight.

Configuration:

appl

V:

3.7.

– Automatic start of averaging when

the Midrics is turned on and a sample
placed on the platform (provided start
conditions are met). Configuration:

appl

V:

3.8.

– Closing application program;

deleting parameters:
The number of measurements remains
active in the memory until you delete it
by pressing the c key, overwrite it or
until you select a different application.

– Restore factory settings. Configuration:

appl

V:

9.1.

A number of measurements are
required, as this forms the basis for
calculation of an average weight.
You can enter the desired number
of measurements, also referred
to as subweighing operations, using
the keypad.

The number you enter is saved until
it is overwritten by another number.
It also remains in memory when
you switch to a different application
program, or switch off the scale.

There are three ways to start the
averaging routine:

– Manual start with preset number of

measurements:
Place the sample on the platform and
press the O key

– Manual start with user-defined number

of measurements:
Place the sample on the platform
and enter the number of weighing
operations using the keypad.
Press the r key to save the number
entered and begin weighing

– Automatic start with preset number

of measurements:
Measurement begins when you place
the first sample on the platform,
provided the start conditions are met.

– Tare function:
1) If you store a tare (weight value) by

pressing the ) key, you can later
enter a tare value manually. The tare
value you enter is added to the stored
tare value.
Setting: menu code 3.25.1
(factory default)

2) A tare value entered manually

overwrites a stored tare value (weight
value). If you enter a tare value manu-
ally, a tare value (weight value) stored
later overwrites the manually entered
value.
Setting: menu code 3.25.2
Operating menu setting:

Appl

V:

3.25

Preparation

§ Switch on the scale: Press e.
§ While all segments are lit,

press the ) key

§ Select the Application menu: Press k

repeatedly until

APPL

is displayed

§ Open the Application menu:

Press the ) key

§ Select the Animal Weighing application:

Press the k key repeatedly until
the desired menu item is displayed and
press ) to open the submenu