Rice Lake Sartorius Midrics 1 and 2 Series - Operating Instructions User Manual

Page 72

Advertising
background image

Example: 1
Checkweighing samples with a target weight of 1250 g and a tolerance range from –10 g to +30 g
Settings (changes in the factory settings required for this example):
Setup: Application: Checkweighing
Setup: PRTPROT (printout): 7.7.x (COM1) (see “Configuration” for options)

1280

6. Enter value for the upper limit

(in this example, 1280 g)

O

Save value for upper limit

7. Weigh samples

p

8. Print the results

Note: If automatic printout
of results is enabled, you do
not need to press the p key;
the results are printed auto-
matically.

Setp

+

1.250 kg

Target

Min

+

1.240 kg

Minimum

Max

+

1.280 kg

Maximum

G#

+

1.256 kg

Gross weight

T

+

0.000 kg

Tare weight

N

+

1.256 kg

Net weight

Lim

+

0.48 %

Percentage of deviation
from target*

W.Diff+

0.006 kg

Absolute deviation from target

-------------------

*

Only in “Tolerance limits”
display mode:
If the weight is lighter than
the target, the display shows:

LL

If the weight is heavier than
the target, the display shows:

HH

72

O

1. Enter the initial target and

tolerance limit values

2. Place a sample equal to the

target weight (in this example,
1250 g) on the platform

O

3. Save target value

1240

4. Enter value for lower limit

(in this example, 1240 g)

O

5. Save value for the lower limit