Rice Lake Sartorius Midrics 1 and 2 Series - Operating Instructions User Manual

Page 54

Advertising
background image

Example:
External calibration and manual adjustment with default weights
(weighing parameters: factory settings)

(

1. Zero the scale.

J

2. Start calibration

(e.g., after calibration prompt:
flashing

WP

symbol).

c.ext.def

is shown for

two seconds.

You are prompted to place
the required weight on the
platform (e.g., 10,000 kg)

3. Position the calibration weight

on the weighing platform

The difference between
measured value and the true
mass is shown with a plus
or minus sign.

Ext.

calibration

Printout is generated,

Nom.

+

10000 g

if adjustment was not

Diff. +

1 g

performed and the procedure
was cancelled by pressing (.

)

4. Start adjustment

(or cancel calibration/adjust-
ment by pressing ().

After adjustment,
the calibration weight value
is displayed.

--------------------

A GMP-compliant printout

24.10.2006

10:15

is generated.

Typ

MW1P1

Ser.no.

12345678

Vers.

1.0103.11.2

BVers.

01-26-06

--------------------
Ext.

calibration

Nom.

+

10000 g

Diff. +

1 g

Ext.

adjustment

Diff. +

0 g

--------------------
24.10.2006

10:15

Name:

--------------------

54