Welch Allyn 300 Series Vital Signs Monitor - User Manual User Manual
Page 21
Directions for Use
Chapter 2 Setup
17
Use the settings menu to select and set the following parameters:
To change a settings parameter:
1.
Select the parameter as indicated above.
2.
Change the value by pressing
or
.
3.
Set the displayed new value either by doing nothing for 10 seconds or by pressing any
button other than
or
. If you press a function button (such as
), the monitor
returns to normal operation with that function (
) activated.
Patient Type
Neonate
Term birth through 28 days, or up to 44 gestational weeks
Pediatric
29 days through 12 years
Adult
13 years and older
Target Pressure
The initial cuff inflation pressure (set individually for each patient type)
Temp Modes
Fahrenheit Predictive
Fahrenheit Monitored
Celsius Predictive
Celsius Monitored
Pulse Tone Volume
From 0 (silent) to 5 (loudest)
ºF
ºF
ºF
ºF M
M
ºC
ºC
ºC
ºC M
M