Directions for Use

Chapter 4 Alarms and Alerts

53

Nonrecoverable Alerts

When the monitor detects a nonrecoverable equipment problem, it does the following

Displays an error code in the SYS window, and shuts off the display to all
other windows

Stops patient monitoring

Stops the pump and opens the air valve

Activates the Nurse Call relay (if connected)

Produces an audible tone

Shuts down as soon as

is pressed or one minute has elapsed

Response

For both recoverable and nonrecoverable equipment alerts, respond as follows:

1.

Press

to immediately silence the alert tone.

2.

Determine what caused the alert and correct the problem.