Chapter 4 Alarms and Alerts
Welch Allyn Vital Signs Monitor 300 Series
6.
Change or accept the MAP high alarm limit as described above (from
step 2
).
7.
Press
to step to the MAP low alarm limit.
8.
Change or accept the MAP low alarm limit.
The range of high and low alarm limits for each vital sign is shown here:
Alarm Limits for Vital Signs
Vital Sign
Low Limit (Default)
High Limit (Default)
Resolution
Systolic (mmHg)
Neonatal
30 - 110 (50)
35 - 115 (100)
5 mmHg (0.5 kPa)
Pediatric
35 - 150 (75)
40 - 155 (145)
5 mmHg (0.5 kPa)
Adult
35 - 250 (75)
40 - 255 (220)
5 mmHg (0.5 kPa)
Diastolic (mmHg)
Neonatal
15 - 95 (30)
20 - 100 (70)
5 mmHg (0.5 kPa)
Pediatric
20 - 120 (35)
25 - 125 (100)
5 mmHg (0.5 kPa)
Adult
25 - 225 (35)
30 - 230 (110)
5 mmHg (0.5 kPa)
Mean Arterial Pressure (mmHg)
Neonatal
15 - 100 (35)
20 - 105 (80)
5 mmHg (0.5 kPa)
Pediatric
20 - 130 (50)
25 - 135 (110)
5 mmHg (0.5 kPa)
Adult
25 - 245 (50)
30 - 250 (120)
5 mmHg (0.5 kPa)
Pulse Rate (beats per minute)
Neonatal
35 - 210 (100)
40 - 215 (200)
5 bpm
Pediatric
35- 210 (50)
40 - 215 (150)
5 bpm
Adult
35 - 210 (50)
40 - 215 (120)
5 bpm
SpO
2
(%)
Neonatal
50 - 98 (85)
51 - 99 (95)
1%
Pediatric
50 - 98 (85)
51 - 99 (---)
1%
Adult
50 - 98 (85)
51 - 99 (---)
1%