Chapter 4 Alarms and Alerts

Welch Allyn Vital Signs Monitor 300 Series

6.

Change or accept the MAP high alarm limit as described above (from

step 2

).

7.

Press

to step to the MAP low alarm limit.

8.

Change or accept the MAP low alarm limit.

The range of high and low alarm limits for each vital sign is shown here:

Alarm Limits for Vital Signs

Vital Sign

Low Limit (Default)

High Limit (Default)

Resolution

Systolic (mmHg)

Neonatal

30 - 110 (50)

35 - 115 (100)

5 mmHg (0.5 kPa)

Pediatric

35 - 150 (75)

40 - 155 (145)

5 mmHg (0.5 kPa)

Adult

35 - 250 (75)

40 - 255 (220)

5 mmHg (0.5 kPa)

Diastolic (mmHg)

Neonatal

15 - 95 (30)

20 - 100 (70)

5 mmHg (0.5 kPa)

Pediatric

20 - 120 (35)

25 - 125 (100)

5 mmHg (0.5 kPa)

Adult

25 - 225 (35)

30 - 230 (110)

5 mmHg (0.5 kPa)

Mean Arterial Pressure (mmHg)

Neonatal

15 - 100 (35)

20 - 105 (80)

5 mmHg (0.5 kPa)

Pediatric

20 - 130 (50)

25 - 135 (110)

5 mmHg (0.5 kPa)

Adult

25 - 245 (50)

30 - 250 (120)

5 mmHg (0.5 kPa)

Pulse Rate (beats per minute)

Neonatal

35 - 210 (100)

40 - 215 (200)

5 bpm

Pediatric

35- 210 (50)

40 - 215 (150)

5 bpm

Adult

35 - 210 (50)

40 - 215 (120)

5 bpm

SpO

2

(%)

Neonatal

50 - 98 (85)

51 - 99 (95)

1%

Pediatric

50 - 98 (85)

51 - 99 (---)

1%

Adult

50 - 98 (85)

51 - 99 (---)

1%