Chapter 3 Patient Monitoring

Welch Allyn Vital Signs Monitor 300 Series

Monitored Temperature Measurement

The procedures for monitored and predictive temperature measurements are the same,
with the following exceptions:

For monitored measurements:

The monitor must be set to take a monitored temperature. (See

“Changing

Temperature Type and Measurement Units”

on page 25.)

The monitor displays the temperature continuously.

The measurement continues until the probe is replaced in the probe holder.

Caution Do not monitor temperature continuously for more than 5 minutes.

Note

Verify that the temperature measurement type is set to monitored. (The letter

, to the right of the temperature display window and below

or

,

is illuminated.)