Directions for Use

Chapter 3 Patient Monitoring

33

If MAP is enabled, MAP results are displayed in the message window.

The measurement display persists for two minutes or until another NIBP cycle is
initiated. If an error is detected, an error tone sounds and an error code appears in
the SYS window.

Automatic NIBP Measurement

Automatic NIBP measurements repeat continuously at programmed intervals.

To set up an automatic NIBP measurement, do the following:

1.

Attach the cuff to the patient’s arm.

2.

Press

to set the measurement interval.

The two dashes (- -) in the message window indicate that automatic measurement is
turned off.

3.

To set an interval, press

or

to cycle through the options, which include - -, ST,

and a range of intervals: 1, 3, 4, 5, 10, 15, 30, 45, 60, 90, 120, and 240 minutes.

Note

If the SpO

2

sensor is attached and generating valid pulse rate data, then the

displayed pulse rate is derived from the SpO

2

sensor reading.

Note

The interval is the time from the beginning of one measurement cycle to the
beginning of the next measurement cycle.

Note

The ST interval selection works differently from the other intervals. For
information on using these settings, please refer to

“STAT Measurement”

on

page 36.

SYS

DIA

SpO

2

%

/min

mmHg

ºC

ºC

M

INTERVAL --