Directions for Use

Chapter 3 Patient Monitoring

35

The automatic NIBP cycles continue until one of the following occurs:

The monitor reaches the 5-minute limit for a STAT measurement. (The current
cycle continues to completion, even if it goes beyond the 5-minute limit.)

The monitor halts because

is pressed.

The monitor halts because of an alarm, alert, or error condition.

The interval code is changed to ‘- -’.

If an error is detected during the measurement, an error tone sounds and an error code
appears in the SYS window.

MAP Measurement

MAP is available for adult, pediatric, and neonatal patients. The monitor is set at the
factory to enable MAP display and alarm limit checking for neonatal patients, and to
disable those functions for adult and pediatric patients.

If MAP is enabled, the monitor displays MAP readings in the message window at the end
of NIBP measurements.

Note

The latest NIBP measurement is displayed until one of the following occurs:

the next NIBP cycle starts

an alarm, alert, or error occurs

the monitor shuts down

SYS

DIA

SpO

2

%

/min

ºC

ºC

kPa

122

MAP 7.5 kPa

6.0

10.7

98

37.2