Monitor specifications, Performance – Welch Allyn 300 Series Vital Signs Monitor - User Manual User Manual

Page 75

Advertising
background image

Directions for Use

Chapter 7 Reference

71

Monitor Specifications

The monitor is an FDA Class II non-critical device and, according to Council Directive 93/
42/EEG, Annex IX, a Class IIB device.

Performance

Characteristic

Specification

General

Recovery from Defibrillation
Discharge

Per IEC 60606-2-30:1999(E) The monitor returns to normal function within 1 minute after
the discharge of a cardiac defibrillator. (Actual recovery time = 0 seconds)

NIBP

Cuff Pressure Range

0 to 300 mmHg (0 to 40 kPa)

Initial Cuff Inflation
Factory Default

Adult

160 mmHg (21.3 kPa)

Pediatric

120 mmHg (16 kPa)

Neonate

90 mmHg (12 kPa)

Blood Pressure Accuracy

Blood pressure measurement meets or exceeds ANSI/AAMI
SP10: 2002 for manual, electronic, or automated
sphygmomanometers.

Blood pressure accuracy is validated for pressure measurement
using the upper arm only, with the patient seated.

Blood pressure is validated against manual auscultatory readings
for adults and children older than 3 years.

Blood pressure is validated against intra-arterial readings for
children 3 years and younger.

Blood Pressure
Determination Time

20 - 45 seconds typical; 165 seconds maximum.

Overpressure Cutoff

295 to 330 mmHg (39.3 to 44 kPa)

Systolic Range

Adult

30 to 260 mmHg (4 to 34.5 kPa)

Pediatric

30 to 160 mmHg (4 to x 21.3 kPa)

Neonate

25 to 120 mmHg (3.3 to 16 kPa)

Diastolic Range

Adult

20 to 235 mmHg (2.7 to 31.3 kPa)

Pediatric

15 to 130 mmHg (2 to 17.3 kPa)

Neonate

10 to 105 mmHg (1.3 to 14 kPa)

MAP

Adult

20 to 255 mmHg (2.7 to 30 kPa)

Pediatric

15 to 140 mmHg (2 to 18.7 kPa)

Neonate

10 to 110 mmHg (1.3 to 14.7 kPa)

Pulse Rate Range

30 to 220 bpm

Pulse Rate Accuracy

± 3 bpm or 3%

Note

NIBP specifications are tested by Welch Allyn using the BIO-TEK BP Pump NIBP
Monitor Tester.

Advertising