Loading a probe cover, Ejecting a used probe cover, Predictive temperature measurement – Welch Allyn 300 Series Vital Signs Monitor - User Manual User Manual
Page 46: ºf ºc ºc
42
Chapter 3 Patient Monitoring
Welch Allyn Vital Signs Monitor 300 Series
Loading a Probe Cover
1.
Holding the probe handle with your thumb and two fingers on the indentations of the
probe handle, withdraw the probe from the probe well.
2.
Insert the probe into a probe cover and press the probe handle down firmly. The
probe handle moves slightly to engage the probe cover.
Ejecting a Used Probe Cover
Do not touch the used probe cover.
1.
Position the probe over an appropriate disposal receptacle.
2.
While holding the probe securely, push the probe cover ejector button (blue or red) to
remove the probe cover into the disposal receptacle.
Predictive Temperature Measurement
To set up for predictive temperatures, please refer to the procedure described in
“Changing Temperature Type and Measurement Units”
To take a predictive temperature, follow these steps:
Note
Verify that the temperature measurement type is set to predictive.
(The display is either
or
; the letter ‘M’ is not illuminated.)
ºF
ºF
ºC
ºC