Directions for Use

Chapter 3 Patient Monitoring

45

Axillary Predictive

When used correctly, the monitor produces an accurate axillary temperature
measurement in less than 15 seconds for adults and in less than 13 seconds for
pediatric patients.

Use Axillary Pediatric (AP) measurements for patients up to 17 years old.

Use Axillary Adult (AA) measurements for patients 18 years old and older.

1.

Remove the temperature probe from the probe holder.

The temperature probe runs a self-test, displaying 188.8 for a few seconds. When it is
ready for use, the temperature window clears, and then OrL appears in the
temperature window.

2.

Press

or

to change the display to AP or AA.

3.

Load a new probe cover by inserting the probe into a probe cover and pressing
the probe handle down firmly. The probe handle moves slightly to engage the
probe cover.

4.

Lift the patient’s arm to fully expose the axilla.

Note

For axillary temperatures, use only the oral probe (blue ejection button) and the
blue probe well.

Caution Use only Welch Allyn probe covers. The use of any other probe cover, or
failing to use a probe cover, can produce measurement errors or inaccuracies.

Note

Be sure that nothing touches the probe tip before you place it in the axillary
measurement site.

Note

Do not allow the probe tip to make contact with the patient until the probe
is placed in the measurement site. Any such contact can cause an
inaccurate reading.