Chapter 3 Patient Monitoring

Welch Allyn Vital Signs Monitor 300 Series

4.

To select the currently displayed interval, press any button other than

,

or

.

Ten seconds after you select an interval, and assuming that safe venous return
pressure (SVRP) has been maintained for at least 30 seconds, the monitor starts the
first automatic NIBP cycle and the following occurs:

The cuff inflates to the default pressure level.

The SYS window dynamically displays the current cuff pressure.

If MAP is enabled, the MAP measurement value alternates with the time display
in the message window.

When the NIBP cycle ends, a tone sounds and the monitor displays the measurement
results, including pulse rate in the

window. (If the SpO

2

sensor is attached to

the patient, the pulse rate is derived from the SpO

2

sensor.)

The measurement display persists until one of the following occurs:

the next cycle begins, if the monitor is still in automatic NIBP mode

two minutes pass

is pressed again

Note

If a MAP alarm occurs, the MAP is displayed steadily in the message window.

Note

If the first cycle does not produce a measurement, the monitor retries the
measurement using a target pressure calculated from the results of the
previous cycle.

SYS

DIA

SpO

2

%

/min

kPa

ºC

ºC

16

.

2

10

.

7

53

20:05:42