Chapter 7 Reference

Welch Allyn Vital Signs Monitor 300 Series

Electrical

Power Requirements

The medical-rated isolation AC power adapter connected to the AC supply must supply 8
volts DC and 0.74 amperes.

Battery

The monitor uses a (supplied) sealed lead-acid, 6V, 4 Ah battery with external
recharge capability.

The battery charges to at least 90% capacity in 12 hours. The AC power adapter, when
connected, simultaneously operates the monitor and charges the battery; it charges the
battery more rapidly if it is not simultaneously operating the monitor.

Using a new, fully charged battery, the monitor can be operated continuously for at least 8
hours of Adult NIBP monitoring, at 3-minute intervals, with simultaneous and continuous
monitoring of temperature and SpO

2

values.

In other words, a new, fully charged battery supports at least 165 Adult NIBP readings at
3-minute intervals while simultaneously and continuously monitoring and displaying
temperature and SpO

2

values.

Leakage Current

For maximum patient electrical isolation, connect a computer to the monitor only when
the monitor is not connected to a patient or when the computer is outside the patient
field running on battery power.

Electrical rating descriptions

Power option

Specification

1

Input: 120VAC 60Hz , 130mA
Output: 8VDC, 750mA

2

Input: 240VAC 50Hz , 65mA
Output: 8VDC, 750mA
Input: 230VAC 60Hz , 65mA
Output: 74VDC, 750mA

4

Input: 240VAC 50Hz , 65mA
Output: 8VDC, 750mA
Input: 230VAC 60Hz , 65mA
Output: 74VDC, 750mA

6

Input: 240VAC 50Hz , 65mA
Output: 8VDC, 750mA
Input: 230VAC 60Hz , 65mA
Output: 74VDC, 750mA

7

Input: 220VAC 50-60Hz , 70mA
Output: 8VDC, 750mA